Standards:
EC60695-2-10/11/12/13,IEC60884-1,GB2099.1,IEC60335-1,GB4706.1,VDE0620
Specification:
1) Volume inside the box: ≥0.5 cubic meters, with glass observation door
2) Illumination in the box: less than 20LUX
3) Glow wire temperature: Normal temperature ~ 1050 °C Adjustable,
4) Thermocouple: Ø1mm nickel-chromium armor (K type) imported thermocouple
5) Temperature tolerance: 500 ~ 750 °C ± 5 °C; 750 ~ 960 °C ± 10 °C
6) The force applied by the glow wire to the sample: 1.0 ± 0.2N, and the penetration depth is limited to 7mm ± 0.5mm
7) Glow wire: Glow wire is a standard size ring made of 4mm diameter nickel/chromium (80/20) wire; the diameter of the
glow wire before it is bent is 4.0 +/- 0.04mm, after bending Reduced to 90%, other sizes are produced according to the error requirements of the standard requirements;
8) The moving speed of the top of the glow wire near and leaving the sample: 10mm / s - 25mm / s;
9) Timer: 0 ~ 99 minutes 99 seconds adjustable, accuracy: ± 1 second hot time (Ta) can be set, light-off duration (Ti) and flame-off time (Te) can be recorded
10) Power supply: AC 220V/50Hz